Utility of a pneumatic system to transport erythrocyte concentrates: Pilot study
Main Article Content
Keywords
Erythrocyte transfusion, Hemoglobins, Hematocrit, Potassium
Abstract
Background: The transport systems of erythrocyte concentrates should be safe and modern. In this paper we analyzed the safety of one pneumatic system to transport erythrocyte concentrates from blood transfusion service to post-surgical therapy service.
Methods: We analyzed 50 erythrocyte concentrates; a sample of 1.5 mL of blood was drawn previously and after delivering the concentrated erythrocyte. The samples were placed in glass tubes to measure temperature, hemoglobin, hematocrit and extracellular potassium concentration. The average and range as univariate analysis and the Mann-Whitney U test as a bivariate analysis were done for statistical analysis.
Results: Statistical differences were not observed in the parameters measured. The temperature was not modified by the pneumatic delivery system. The data indicated no significant differences (p > 0.05) pre and post delivered samples in temperature, hemoglobin concentration, hematocrit level or extracellular potassium concentration. The estimated travel time for each sample was 22.88 seconds.
Conclusions: The transport of erythrocyte concentrates through pneumatic delivery system did not change temperature conditions according to the current regulation. Indeed, the data did not show evidence that the cellular suspension underwent a spontaneous cellular lysis by sending erythrocyte concentrates through a pneumatic delivery system.
References
Fernandes CM, Worster A, Eva K, Hill S, McCallum C. Pneumatic tube delivery system for blood samples reduces turnaround times without affecting sample quality. J Emerg Nurs. 2006;32(2):139-43.
Lancé MD, Marcus MA, van Oerle R, Theunissen HM, Henskens YM. Platelet concentrate transport in pneumatic tube systems. Does it work? Vox Sang. 2012;103(1):79-82.
Sodi R, Darn SM, Stott A. Pneumatic tube system induced haemolysis: Assessing sample type susceptibility to haemolysis. Ann Clin Biochem. 2004; 41(Pt 3):237-40.
Council of Europe Publishing. Guide to the preparation, use and quality assurance of blood components. Seventeenth edition. Strasbourg Cedex, France: Council of Europe Publishing; 2013.
Norma Oficial Mexicana NOM-003-SSA2-2012, para el uso de sangre y sus componentes con fines terapéuticos. Diario Oficial de la Federación del 26/10/2012. Disponible en http://www.dof.gob.mx/nota_detalle.php?codigo=5275587&fecha=26/10/ 2012
Stair TO, Howell JM, Fitzgerald DJ, Bailey SC, Bastasch MD. Hemolysis of blood specimens transported from ED to laboratory by pneumatic tube. Am J Emerg Med. 1995;13(4):484-8.
Ellis G. An episode of increased hemolysis due to a defective pneumatic air tube delivery system. Clin Biochem. 2009;42(12):1265-9.
Streichert T, Otto B, Schnabel C, Nordholt G, Haddad M, Maric M, Petersmann A, et al. Determination of hemolysis thresholds by the use of data loggers in pneumatic tube systems. Clin Chem. 2011;57(10): 1390-7.